Título: Estabelecimento de uma curva de calibração para o equipamento bactcount para monitoramento da qualidade do leite cru refrigerado
Autor: JATOBÁ, Raquel Bezerra de 
Primeiro orientador: BARBOSA, Severino Benone Paes
Primeiro coorientador: BATISTA, Ângela Maria Vieira
Segundo coorientador: MONARDES, Humberto Gonzalo
Primeiro membro da banca: SENA, Maria José de
Segundo membro da banca: PIRES, Edleide Maria de Freitas
Resumo: O experimento foi conduzido no Laboratório de Qualidade do Leite, Programa de Gerenciamento de Rebanhos Leiteiros do Nordeste ¿ PROGENE, situado no Departamento de Zootecnia da Universidade Federal Rural de Pernambuco. Objetivou-se estabelecer uma curva de calibração para o equipamento Bactocount IBC, que realiza contagem individual de bactérias (CIB) no leite, o qual deverá expressar os resultados em unidade formadora de colônia (UFC), preconizado pela IN 51 (MAPA, 2002). A UFC é um dos parâmetros de monitoramento da qualidade do leite cru refrigerado de maior importância, uma vez que, a microbiota existente pode servir como veículo para doenças transmissíveis ao homem. Foram coletadas 169 amostras de leite de tanques de expansão e de latões de quatro Estados da região Nordeste e, em seguida, acondicionadas em recipientes estéreis com capacidade para 50 mL e conservadas a 4ºC (± 1) até a realização das análises, que não ultrapassou 48 horas. Após a análise no equipamento Bactocount IBC, as amostras foram encaminhadas para determinação de UFC utilizando-se o método de referência com placas Petrifilm® AC (AOAC, 2002;BRASIL, 2003). A curva de calibração foi construída com dados obtidos do contador individual de bactérias (CIB) comparando-os com os resultados do método de referência em UFC. Foi aplicada uma análise de correlação e posteriormente de regressão linear simples. A correlação entre CIB e UFC observada foi de 0,98 e a equação obtida foi a seguinte: logUFC = - 0,6324 + 1,102 * logIBC
Abstract: The experiment was conducted at PROGENE laboratory in the Rural Federal University of Pernambuco. The aim of this paper was to develop a calibration curve for the Bactocount IBC equipment. This equipment perform the individual counting of bacteria in milk, but it is necessary to express the results of this counting in colonies formed per unit (CFU). The CFU is one of the most important parameters indicating the quality of the raw milk because milk may carry a variety of microrganisms that can infect humans. 169 samples were collected from bulk milk from four states of Northeast region of Brazil and stored in a sterile device at 4ºC and brought to the laboratory. After the analysis in Bactocount equipment, the analysis by the CFU method was performed by using the standard count in Petri dish. A calibration curve was constructed with data from Bactocount equipment compared to the CFU method. It was applied a correlation analysis and a linear regression. The correlation between Bactocoun and CFU was 0,98 and equation generated was log(CFU) = - 0,6324 + 1,102 * log(IBC).
Palavras-chave: Citometria de fluxo.
Contagem bacteriana
Qualidade do leite
Bacterial count
Flow cytometry
Quality of milk
Área(s) do CNPq: CIENCIAS AGRARIAS::ZOOTECNIA
Idioma: por
País: Brasil
Instituição: Universidade Federal Rural de Pernambuco
Sigla da instituição: UFRPE
Departamento: Departamento de Zootecnia
Programa: Programa de Pós-Graduação em Zootecnia
Citação: JATOBÁ, Raquel Bezerra de. Estabelecimento de uma curva de calibração para o equipamento bactcount para monitoramento da qualidade do leite cru refrigerado. 2009. 48 f. Dissertação (Programa de Pós-Graduação em Zootecnia) - Universidade Federal Rural de Pernambuco, Recife.
Tipo de acesso: Acesso Aberto
URI: http://www.tede2.ufrpe.br:8080/tede2/handle/tede2/6871
Data de defesa: 18-Fev-2009
Aparece nas coleções:Mestrado em Zootecnia

Arquivos associados a este item:
Arquivo Descrição TamanhoFormato 
Raquel Bezerra Jatoba.pdfDocumento principal420,12 kBAdobe PDFBaixar/Abrir Pré-Visualizar


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.